Resumen
The method specified is based on drawing continuously, at a constant flow rate, an air sample through a particle filter before it enters the chemiluminescence analyser. It then flows into a reaction chamber where it is mixed with an excess flow of ethylene. Ozone and ethylene react instantaneously to produce light in the visible region (maximum at about 400 nm). The emitted light intensity is proportional to the concentration of ozone in the air sample and is measured by a photomultiplier tube. Applies to determination of the mass concentration of ozone between 2 µg/m^3 and 10 mg/m^3 at the reference conditions.
Informaciones generales
-
Estado: PublicadoFecha de publicación: 1993-04Etapa: Norma Internacional confirmada [90.93]
-
Edición: 1Número de páginas: 9
-
Comité Técnico :ISO/TC 146/SC 3ICS :13.040.20
- RSS actualizaciones
Modificaciones
Las enmiendas se emiten cuando se decide que puede ser necesario agregar nuevo material a un documento de normalización existente. También pueden incluir correcciones editoriales o técnicas que deban aplicarse al documento existente.
Amendment 1
Consensus value for the ozone absorption cross-section at room temperature at the mercury-line wavelength (253,65 nm air)
Edición 2024
Ciclo de vida
-
Ahora
-
00
Preliminar
-
10
Propuesta
-
20
Preparación
-
30
Comité
-
40
Consulta
-
50
Aprobación
-
60
Publicación
-
90
Revisión
-
95
Retirada
Modificaciones
Aportan contenido adicional; disponible para su compra; no incluidas en el texto de la norma existente.PublicadoISO 10313:1993/Amd 1:2024
-
00